F141x-Q Dual QCM Standard Dewpoint Transmitter

Product Advantages:Ultra-low dewpoint measurement Measurement Range:-120 … 0 °Ctd

Leaflet

QCM sensor technology is a revolutionary humidity measurement technology. Combined with the newly developed moisture sensitive material, it has excellent signal sensitivity in ultra-low humidity conditions and can measure humidity down to ppb level stably Patented dual QCM sensors, one for humidity signal and one for pollution signal, automatically compensate for the drift caused by pollution which is common in QCM humidity measurement technology Innovative temperature compensation algorithm and multi-point temperature-compensated calibration, greatly reduces temperature-dependent dewpoint drift and ensuring high-precision dewpoint measurement over a wide temperature range. Combined with the ultra-high stability of QCM sensor technology, recalibration intervals can be extended to more than two years Innovative temperature compensation algorithm and multi-point temperature compensation calibration, greatly improving the sensor’s temperature drift and ensuring high-precision dewpoint measurement over a wide temperature range

  • Revolutionary FixInst-Q QCM humidity sensitive material, lower limit of dewpoint down to -120 °Ctd, ppb level of humidity
  • Application in dewpoint < -80 °Ctd: High-purity industrial gas, clean gas
  • Patented dual QCM sensors, one for humidity signal and one for pollution signal, automatically compensate for the drift caused by pollution
  • Accurate measurements up to ±2 °Ctd
  • Ultra-fast response time and outstanding long-term stability
  • Multi-point temperature compensation calibration
  • High resistance to electrical disturbance
  • IP65 protection class, providing good protection even in harsh environments
  • Provides comprehensive sensor setup, data transfer, software upgrades, and maintenance via Modbus RTU (RS485) interface and powerful service APP
Ideal for ultra-low dew point monitoring in high-purity industrial gases, clean gas systems, and semiconductor manufacturing. Ensures precise moisture measurement, gas purity, and stable process performance.

Measuring Range

Dewpoint
F141B-110 … 0 °Ctd
F141C-120 … 0 °Ctd
Temperature-40 … +100 °C

Output

Analog Output (Customized)4 … 20 mA (3-wire)
Analog Resolution0.002 mA
Analog Drift0.01% of span/°C
Digital OutputMax. 500 ohm
Relay OutputModbus RTU (RS485)
Connector5pin M12, Female

Accuracy

Dewpoint (Air or Nitrogen)
+20 … -80 °Ctd±2 °Ctd
-80 … -120 °Ctd±3 °Ctd
Temperature (Customized)
0 … +50 °C±0.3 °C (Standard)
-40 … 0 °C & +50 … +100 °C±0.5 °C (Standard)

Operating Environment

Environment Temperature-20 … +70 °C
Storage Temperature-30 … +80 °C
Relative Humidity0 … 95 %RH
Sample Gas Flow Rate> 1 L/min
Pressure0 … 5 MPa(a)

Response Time

Dewpoint: 63% [90%], Reference: 20 °C, 1bar(a), 4L/min
-80 → +30 °Ctd20 sec [40 sec]
+30 → -70 °Ctd5 min [20 min]

Other

Process ConnectionISO G1/2″ thread (Standard)
3/4″ – 16 UNF thread (Customized)
5/8″ – 18 UNF thread (Customized)
Protection CodeIP65
Housing MaterialSUS304
Sensor FilterStainless steel mesh filter
(Filtration class 70 um)
EMCCompliant with IEC 61326-1

Power

Measuring State 10 … 30 VDC Max 1.5W @ 24VDC
Leaflet_F141x_Dewpoint-Meter_YUA_EN_2024-V3.01.pdf396 KB
Download ↓
QG_General-Dewpoint-Meter_YUA_EN_2024-V1.03-2.pdf370 KB
Download ↓

CONTACT

Please input your name
Please input Email
Please input Tel
Please input Company
Please input Business
Please input needs
Please fill in the above information accurately. We will contact you as soon as possible!
×

Write to Us

Tell us about your application and measurement requirements.
Our technical team will recommend the right solution.