F108x FIXINST-A New-generation Al₂O₃ Standard Dewpoint Transmitter

Product Advantages:Fast Response Measurement Range:-110 … +20 °Ctd

Leaflet

When the dewpoint of the measuring medium is below -40 °Ctd, the aluminum oxide sensor is the most accurate, responsive and long-term stable in comparison with other types of sensors. Meanwhile, it has excellent anti-pollution ability Revolutionary auto-calibration circuit design automatically compen sates for circuit drift caused by temperature, contamination, and aging, providing long-term stability and high accuracy measurements Innovative temperature compensation algorithm and multi-point temperature compensation calibration, greatly improving the sensor’s temperature drift and ensuring high-precision dewpoint measurement over a wide temperature range

  • FixInst new generation ultra-fast response aluminum oxide sensor technology
  • Application in dewpoint < -60 °Ctd: Desiccant dryers, nitrogen generators, industrial gases
  • 10+ points of dewpoint calibration and multi-point temperature compensation provide accurate measurements up to ±2 °Ctd (Refer to “Technical Data”)
  • Ultra-fast response time and outstanding long-term stability
  • Anti-condensation, resistant to particulate contamination, oil vapor and most chemicals
  • Excellent immunity to interference
  • IP65 protection class, providing good protection even in harsh environments
  • Provides comprehensive sensor setup, data transfer, software upgrades, and maintenance via Modbus RTU (RS485) interface and powerful service APP
  • Calibration is valid for two years, reducing maintenance

Ideal for low dew point monitoring in compressed air systems, desiccant dryers, nitrogen generators, and industrial gas pipelines. Helps maintain gas dryness and prevent moisture-related equipment damage.

Measuring Range

Dewpoint
F108A-80 … +20 °Ctd
F108B-110 … +20 °Ctd
Temperature-40 … +100 °C

Output

Analog Output (Customized)4 … 20 mA (3-wire)
Analog Resolution0.002 mA
Analog Drift0.01% of span/°C
Digital OutputMax. 500 ohm
Relay OutputModbus RTU (RS485)
Connector5pin M8, Female

Accuracy

Dewpoint (Air or Nitrogen)
+20 … -60 °Ctd±2 °Ctd
-60 … -100 °Ctd±3 °Ctd
Temperature (Customized)
0 … +50 °C±0.3 °C (Standard)
-40 … 0 °C & +50 … +100 °C±0.5 °C (Standard)

Operating Environment

Environment Temperature-30 … +70 °C
Storage Temperature-40 … +80 °C
Relative Humidity0 … 95 %RH
Sample Gas Flow Rate> 1 L/min
Pressure0 … 5 MPa(a)
0 … 35 MPa(a) (Option)

Response Time

Dewpoint: 63% [90%], Reference: 20 °C, 1bar(a), 4L/min
-50 → +20 °Ctd20 sec [40 sec]
+20 → -50 °Ctd1 min [3 min]

Other

Process ConnectionISO G1/2″ thread (Standard)
3/4″ – 16 UNF thread (Customized)
5/8″ – 18 UNF thread (Customized)
Protection CodeIP65
Housing MaterialSUS304
Sensor FilterStainless steel mesh filter
(Filtration class 70 um)
EMCCompliant with IEC 61326-1

Power

Measuring State 10 … 30 VDC Max 1.5W @ 24VDC
Leaflet_F108x_Dewpoint-Meter_YUA_EN_2024-V3.01.pdf430 KB
Download ↓
QG_General-Dewpoint-Meter_YUA_EN_2024-V1.03-1.pdf370 KB
Download ↓
IM_F108x_Dewpoint-Meter_YUA_EN_V1.03.pdf603 KB
Download ↓

CONTACT

Please input your name